Blower Box With Fan & Thermal Shutoff

At the heart of our Portable Fume Hoods, Downdraft Tables, and Workstations is the Blower Box. This is the work horse of the system and, largely, the reason our systems are so effective. The Blower Box is generally a replacement item as the Blower Box comes standard with our Portable Fume Hoods, Downdraft Tables, and Workstations. 

Item Product Price QTY
52491 Blower Box w/Fan & Thermal Shutoff Login

Hazard Technology is part of Stoelting Co. For technical support or service contact: info@StoeltingCo.com or 800-860-9775


Simple and Effective. At the heart of our Portable Fume Hoods, Downdraft Tables, and Workstations is the Blower Box. This is the work horse of the system and, largely, the reason our systems are so effective. The Blower Box is generally a replacement item as the Blower Box comes standard with our Portable Fume Hoods, Downdraft Tables, and Workstations. 

Please Note: Upon arrival, Blowers do NOT contain a Filter. A Filter MUST be installed for proper function.

Without a Filter, the system offers no protection from hazerdous fumes, vapors, or particles.

Filters should be changed regularly and according to your specific use and needs. Even without use, once a filter has been installed, it should be changed at least once per year. Failure to replace the filter increases the risk that your hood will not provide appropriate protection from hazardous materials.

Filter options include:

Carbon (Charcoal) - rids the work area of hazardous fumes and vapors.

HEPA - used for containment of particles - toxic powders, asbestos, lead, latent fingerprint powder, etc.

Potassium Permanganate - safely absorbs and neutralizes formaldehyde and paraformaldehyde fumes.

Polyester – generally used for dust prevention and as a pre-filter in combination with HEPA filters

*Filters may be stacked for mixed use applications*

You can find Filters for Portable Fume Hoods, Downdraft Tables and Workstations Here

 

Installation of the Blower Box with Portable Fume Hoods, Downdraft Tables and Workstations

1.     Remove Packaging by cutting away outside boxes

2.     Open Blower Box

3.     Remove Wingnut Washers and Filter Holders from Threaded Rod

4.     Place Filter in Plenum (Filter Gasket facing down)

5.     Be certain the Filter Gasket is between guides

6.     If using two Filters place Dust Filter in first and then Vapor Filter on top

7.     Replace Filter Holders and Wingnut Washers

8.     Tighten Wingnut Washers to depress Filter Holders

9.     It is necessary to depress Filter Gaskets to create an appropriate seal

Important – Do Not touch HEPA Filter media

10.  Close Blower Box

Scavengers Users Only

11.  Attach hoses as required

12.  Attach nozzles as required

13.  Follow steps 17-21

Portable Fume Hood Users Only

14.  Place Blower on top of Acrylic Enclosure

15.  Make sure Blower(s) is placed on top of enclosure correctly

16.  See attached diagram for correct placement

All Users

17.  Plug Blower(s) into 110-volt grounded outlet

18.  Turn Blower(s) on

19.  Adjust cover to working height or cut cover as needed

20.  You are now ready to use your Portable Fume Hood